Description

Features:
- Automatic adjustment of ventilation parameters and alerts borders with due account for patient’s weight;
- High sensitive correction of the ventilation in case of independent breathing attempts of the patient
- Automatic compensation of the breathing circuit’s

compliance influence to the Inspiration volume

Automatic selection of pressure increase optimum rate

(Pramp - AUTO)

Sensitive smart trigger does not miss real breathing attempts and minimizes the probability of false alarm even under outflow

Auto selection of the optimal moment for switching to exhalation when spontaneous inspiration (Pramp -AUTO)

Apnea detection

Option for individual setting of apnea ventilation parameters

Adjustment of ventilation pattern to the individual parameters of lungs

Automatic data recalculation to BTPS conditions

Automatic outflow compensation.

The graphic interface provides doctor with logic and convenient complex for over patching of parameter and its further monitoring.

The ventilator has detailed digital and graphic monitoring of all main parameters.

The ventilator is equipped by touch screen display. 12´´ flat-panel liquid-crystal color display represents diagrams and numerical data in real-time operation mode and provides management of all parameters and functions of ventilator.

Monitoring parameters:
Exhalated Tidal Volume (Vte) Peak Inspiratory pressure (Ppeak)
Inspirated Tidal Volume (Vti) Mean Airway Pressure (Pmean)
Minute ventilation (MV, MVspont) Plateau pressure (Pplat)
Breath rate (f, fspont) Positive End Expiratory Pressure (PEEP)
Inspiratory time (Ti) FiO2
Expiratory time (Te) Compliance (Cstat, Cdyn)
I :Е Ratio Resistance (Raw)
SpO2

EtO2

EtC02, FiC02
Trends up to 7 days

Graphic systemgrafik

displays the following graphics:
curves (Flow, Pressure, Volume, Capnogramm, SpO2);
loops (Flow/Volume, Pressure/Volume)
trends (up to 24 hours)
Gas sources
Compressed gas distribution system, oxygen cylinder, concentrator (O2)

compressor, coming with the ventilator (air)

Perfomance specifications:
Breath rate (f) 1-150 bpm
Tidal volume (VT) 20-2500 ml
Inhalation Peak flow (Peak flow) Not less than 240 l/min
Pressure support(PS) 0-1000 смН2О
Outflow compensation. On/of
Manual control of inhalation
Inspiratory pause
Expiratory pause
FiO2 increasing (100% O2 at 2 min)
Display on hinges
Block of touch panel