The device provides support for a constant pressure waveform at all breathing frequencies, has an alarm system, has a built-in oxygen analyzer with a digital indicator and the ability to filter exhaled gases.


Modes of ventilation Off / Alarm Test, CPAP, CMV, PTV, SIMV

BRP (respiratory rate) 1 - 125 (126 - 250 with option Z0004)

Maximum inspiratory time, sec 0.1 - 3.0, (0.01 - 0.3 with option Z0004)

Inspiratory / expiratory ratio (I: E) from 9.9: 1 to 1:99

The flow pressure is weak or strong, set by switch

Manual breath gives a single breath in CPAP, CMV, PTV, SIMV mode

PTV sensitivity allowed maximum

Oxygen concentration 21 - 100% O2 ± 3%

Pressure mode switch max - average - minimum

Pressure in CPAP mode, mbar 0 - 15

Inspiratory pressure, mbar 0 - 60

Disable alarm for 60 seconds

Resetting alarms resets all alarms (excluding system errors)

Light indicators:

Network is green, network is on

System error red signal, error in the main processor

Backup trigger green signal, indicates the machine is breathing

Digital:

Breath rate monitor breaths per minute

Monitor inspiratory time inspiratory time

Inspiration / expiration ratio (I: E) inspiratory / expiratory value

Oxygen concentration,% from 21 to 100

Pressure, mbar from 0 to 62

Calibration:

Pressure calibrated, mbar 0 to 60 plus PEEP level

Anxiety:

CPAP visual and intermittent beeps

PIP / cycle error / low visual and intermittent beeps

Loss of main power battery power, beep

Loss of air or oxygen supply pneumatic, sounds from a blender

System error visual and intermittent beeps

Equipment, size and weight:

Inlet pressure of air and oxygen, bar 3 - 5

Voltage 100-120 V, 50/60 Hz; 220-240 V, 50/60 Hz

Power, W 20

Fuses 100-120 V = T500 mA 220-240 V = T200 mA

Protection Class I Type B

Dimensions (machine only), cm 37 (length) x 31 (height) x 32 (width)

Stand height, cm 137