The device of artificial ventilation of the lungs for newborns SLE 5000 without HFO
The SLE 5000 medical ventilator is intended for full or partial prosthetics of the respiratory function of newborns weighing 300 g or less with respiratory failure. It is used in resuscitation departments of medical institutions of obstetric aid.
Part number:
Supplier:
Shvabe Description
Modes of ventilation:
- CMV - forced ventilation of the lungs.
- PTV - lung ventilation initiated by the patient.
- SIMV - synchronized intermittent forced ventilation of the lungs.
- CPAP - breathing support with constant positive pressure, manual inhalation, non-invasive ventilation.
- TTV + - ensuring a stable expiratory volume at the lowest possible pressure at a set time for inspiration Ti. The main difference from TTV - the algorithm provides a change in pressure, but not the time of inspiration.
- PEEP - positive expiratory pressure.
- PSV - pressure assisted ventilation.
Benefits:
- Ability to combine multiple ventilation modes (CMV + TTV, SIMV + TTV + PSV).
- Display of data on the built-in color LCD display with the Russified menu.
- Ability to synchronize modes and the implementation of the principles of "protective" ventilation of the lungs. - Support constant pressure at all frequencies.
- Filtration of exhaled gases.
- The possibility of using gas mixtures of nitric oxide.
- Emergency 3-level alarm system with light indication.
- Control of the mechanical properties of the lungs.
- Valveless technology, eliminating the pneumatic effect.
- Use a single circuit for all modes.
- High-frequency mode (optional).
Ventilation modes traditional CPAP / PTV / PSV
Breath time
0.1 - 3.0 seconds
CPAP pressure
0 - 20 mbar
Inhalation pressure
0 - 65 mbar
Target volume
2 - 200 ml
FiO2
21% - 100%
CMV / SIMV
BPM (respiration rate) 1 - 150 vd / min
Inspiration / expiration ratio I: E:
11.2: 1 - 1: 600
Breath time
0.1 - 3.0 seconds
PEEP pressure
0 - 20 mbar
Inhalation pressure
0 - 65 mbar
Target volume
2 - 200 ml
FiO2
21% - 100%
Traditional ventilation and combined modes
Patient leakage
0 - 50% (Resolution - 5%, average for the last 5 breaths)
Respiratory rate total
0 - 150 dpi / min
Dynamic compliance
0 - 100 ml / mbar (Resolution - 1 ml / mbar)
C20 / C
Resolution 0.1
Measurement time
2 ms
Resistance
0 - 1000 mbar.s / l
Flow trigger sensitivity
0.2 - 10 l / min
Oxygen concentration
Range 21% - 100% (resolution 1%)
Pressure
Real-time pressure measurement Resolution 1 mbar
Measurement time
2 ms
Peak pressure
0 - 175 mbar (resolution 1 mbar)
PEEP pressure
0 - 175 mbar (resolution 1 mbar)
Mean Pressure
-175 -175 (resolution 1 mbar)
Manufacturer JSC "PO" UOMZ "
- CMV - forced ventilation of the lungs.
- PTV - lung ventilation initiated by the patient.
- SIMV - synchronized intermittent forced ventilation of the lungs.
- CPAP - breathing support with constant positive pressure, manual inhalation, non-invasive ventilation.
- TTV + - ensuring a stable expiratory volume at the lowest possible pressure at a set time for inspiration Ti. The main difference from TTV - the algorithm provides a change in pressure, but not the time of inspiration.
- PEEP - positive expiratory pressure.
- PSV - pressure assisted ventilation.
Benefits:
- Ability to combine multiple ventilation modes (CMV + TTV, SIMV + TTV + PSV).
- Display of data on the built-in color LCD display with the Russified menu.
- Ability to synchronize modes and the implementation of the principles of "protective" ventilation of the lungs. - Support constant pressure at all frequencies.
- Filtration of exhaled gases.
- The possibility of using gas mixtures of nitric oxide.
- Emergency 3-level alarm system with light indication.
- Control of the mechanical properties of the lungs.
- Valveless technology, eliminating the pneumatic effect.
- Use a single circuit for all modes.
- High-frequency mode (optional).
Ventilation modes traditional CPAP / PTV / PSV
Breath time
0.1 - 3.0 seconds
CPAP pressure
0 - 20 mbar
Inhalation pressure
0 - 65 mbar
Target volume
2 - 200 ml
FiO2
21% - 100%
CMV / SIMV
BPM (respiration rate) 1 - 150 vd / min
Inspiration / expiration ratio I: E:
11.2: 1 - 1: 600
Breath time
0.1 - 3.0 seconds
PEEP pressure
0 - 20 mbar
Inhalation pressure
0 - 65 mbar
Target volume
2 - 200 ml
FiO2
21% - 100%
Traditional ventilation and combined modes
Patient leakage
0 - 50% (Resolution - 5%, average for the last 5 breaths)
Respiratory rate total
0 - 150 dpi / min
Dynamic compliance
0 - 100 ml / mbar (Resolution - 1 ml / mbar)
C20 / C
Resolution 0.1
Measurement time
2 ms
Resistance
0 - 1000 mbar.s / l
Flow trigger sensitivity
0.2 - 10 l / min
Oxygen concentration
Range 21% - 100% (resolution 1%)
Pressure
Real-time pressure measurement Resolution 1 mbar
Measurement time
2 ms
Peak pressure
0 - 175 mbar (resolution 1 mbar)
PEEP pressure
0 - 175 mbar (resolution 1 mbar)
Mean Pressure
-175 -175 (resolution 1 mbar)
Manufacturer JSC "PO" UOMZ "